RegulationIllustrative imageIn Vitro Diagnostic Regulation (IVDR)
Sets the risk-based framework for in-vitro diagnostic devices, with four classes, performance evaluation requirements and a substantially expanded role for notified bodies.
IVDs are classified into classes A to D according to the risk to the individual and to public health, as defined in Annex VIII. Class D covers the highest-risk assays, including blood grouping and detection of lethal transmissible agents, and requires notified body assessment plus reference laboratory involvement.
The regulation requires a performance evaluation that demonstrates scientific validity, analytical performance and clinical performance. Manufacturers must document the intended use precisely, since a change in claimed use can alter both the class and the evidence required.
Post-market performance follow-up is a distinct obligation for IVDs, and the performance evaluation report must be updated with data gathered after placing the device on the market. Companion diagnostics additionally require consultation with a medicinal product authority.
Devices manufactured and used within a single health institution may fall outside the regulation where the conditions in Article 5(5) are met, including appropriate quality management and a justification that no equivalent CE-marked device is available on the market.